Connection: Can you comment on the burden in the laboratory, if one changes from a current ER/PR assay to a standardised ER/PR assay? Important decisions in standardising these assays are the choice of fixation, processing, pre-treatment, primary antibody and visualisation method. These choices can be helped by consulting external quality assurance schemes (4,5), which provide details of which methods have worked well. As mentioned earlier to the answer to the first question, a weak positive external control is important. It can take a while to find a suitable control as only a few percent of tumours are weakly positive. There is obviously work involved in ensuring that a new method works satisfactorily, but being a member of an external quality assurance scheme can provide helpful support in both establishing and maintaining a high quality service. I am fortunate to work in a department with an excellent immunohistochemistry laboratory that is able to do this work. An alternative is to use a commercially available standardised kit. Such kits mean that one does not have to make the choices mentioned above, but one must follow the instructions carefully, and one still has to ensure that the method is working well. Connection: What is your opinion regarding the use of ER/PR pharmDxTM Kit as an aid in the prognosis and management of breast cancer? How do you compare this kit to your own method both in relation to clinical value and ease-of-use? Can kits standardise ER/PR - since there are so many methods available? We compared the staining for oestrogen and progesterone receptor using the pharmDxTM Kits and our standard methods, which have been developed over a number of years. The results were very similar, but the staining was slightly stronger with the pharmDxTM method for oestrogen receptor, which led to a slightly higher positive rate (1% increase). Both methods produce satisfactory results for clinical use. The immunohistochemical methods are both straightforward to use. The pharmDxTM method has fewer steps. However, it does require pre-treatment with a specific Pascal pressure cooker, which can take a limited number of slides per batch and is slower to use than microwave pre-treatment. Developing in-house methods requires more work to validate and to ensure standardisation, but is cheaper. One suggestion that we would have to improve the pharmDxTM Kits would be the addition of a weakly positive control to the negative and strongly positive ones. Tests that have evolved over time, such as immunohistochemistry for hormone receptors, tend to have a variety of methodologies and as a consequence are more difficult to standardise. Nevertheless, the key principles of validation and standardisation of therapeutic assays remain paramount. Laboratories with limited expertise in assay development should consider the use of standardised kits. Such kits can improve the immunohistochemical part of the assay, but the questions of scoring discussed above still remain. Acknowledgements: I would like to acknowledge John Ronan who runs our excellent immunohistochemistry laboratory and other colleagues in the department who have also contributed to the audits mentioned above, including Claire Paish, Shaffiq Gill, Zsolt Hodi and Ian Ellis. References 1. Hodi Z, Chakrabarti J, Lee AHS, et al. The reli ability of assessment of oestrogen receptor expression on needle core biopsy specimens of invasive carcinomas of the breast. J Clin Pathol 2007;60:299-302. 2. Collins LC, Botero ML, Schnitt SJ. Bimodal frequency distribution of estrogen receptor immunohistochemical staining results in breast cancer: an analysis of 825 cases. Am J ClinPathol 2005;123:16-20. 3. Nadji M, Gomez-Fernandez C, Ganjei-Azar P, et al. Immunohistochemistry of estrogen and progesterone receptors reconsidered: experience with 5,993 breast cancers. Am J Clin Pathol 2005;123:21-27. 4. http://www.uknequasicc.ucl.ac.uk 5. http://www.nordiqc.org 6. Goldstein NS, Ferkowicz M, Odish E, et al. Minimum formalin fixation time for consistent estrogen receptor immunohistochemical staining of invasive breast carcinoma. Am J Clin Pathol 2003;120:86-92. 7. Barnes DM, Harris WH, Smith P, Millis RR, Rubens RD. Immunohistochemical determination of oestrogen receptor: comparison of different methods of assessment of staining and correlation with clinical outcome of breast cancer patients. Br J Cancer 1996; 74: 1445-1451 8. Elledge RM, Green S, Pugh R, et al. Estrogen receptor (ER) and progesterone receptor (PgR), by ligand-binding assay compared with ER, PgR and pS2, by immuno-histochemistry in predicting response to tamoxifen in metastatic breast cancer: a Southwest Oncology Group Study. Int J Cancer 2000; 89: 111-117 9. Harvey JM, Clark GM, Osborne CK, et al. Estrogen receptor status by immunohistochemistry is superior to the ligand-binding assay for predicting response to endocrine therapy in breast cancer. J Clin Oncol 1999:17;1474-81 33 | Connection 2008
April 2008, VOL 11 Connection IHC STANDARDIZATION AROUND THE WORLD In this issue Q&A with Patho
Contents APRIL 2008, VOL 11 2 3 4 Editorial George L. Kumar, PhD Featured Laboratory The Osamura
Editorial Immunohistochemistry (IHC) Standardization Around the World George L. Kumar, PhD Managing
Featured Laboratory The Osamura Laboratory Laboratory headed by Prof. Robert Yoshiyuki Osamura, MD,
Q&A Ask the Experts: On Immunohistochemistry (IHC) Standardization Professor Leong is Medical Di
time interval between removal of tissue and immersion in fixative, the temperature of tissue storage
Connection: How would you like to standardize IHC in Australia? As discussed above, standardization
Connection: Antigen retrieval is essential in immunohistochemistry (IHC) in order to restore epitope
Connection: Is a universal image analysis system feasible? No, not until we standardize the all impo
Q&A Fernando Soares, MD, PhD University of São Paulo, São Paulo, Brazil Dr. Fernando Soares is F
Connection: In your opinion, what is the biggest hurdle for standardization? Probably education in l
There are no standards in Latin America. We always follow the manufacturers protocol during our firs
Q&A Bryan R. Hewlett, ART, MLT Consultant Technologist to the Anatomic Pathology EQA program of
There is no standardization of AR steps circumstances, it would be impossible toand, undera the pre
Connection: How would you rate European (UK NEQAS, NordiQC) and US IHC standards to Canadian IHC sta
Q&A Prof. Chen Jie Prof. Cui Quancai Peking Union Mediacl College Hospital, Beijing, China Prof.
Connection: According to Goldstein et al. Appl Immunohistochem Mol Morphol 2007;15:124133 Immunohis
Connection: Is it true that a particular histology feature may be better demonstrated by other fixat
Connection: What constitutes standardization of image analysis as applied to immunohistochemistry (I
Q&A Dr. Tanuja Manjanath Shet Dr. Vani Parmar Tata Memorial Hospital, Mumbai, India Tanuja Manj
... the biggest hurdle in India is suboptimal fixation and processing of tissues. Though I agree th
Connection: Is it true that a particular histology feature may be better demonstrated by other fixat
Connection: Can you comment on the internal and external quality control (EQC) procedures followed i
Q&A Prof. Robert Yoshiyuki Osamura Department of Pathology, Tokai University School of Medicine
Connection: In your opinion, what is the biggest hurdle for standardization? The biggest hurdle for
the standardization of image ... appropriate for pre-screeninganalysis is of the staining quality.
Connection: Why is standardization of image analysis in diagnostic pathology important? Because the
Q&A James F. Happel, DLM (ASCP), HTL Technical Director of Surgical Pathology, Research and Deve
Connection: United Kingdom National External Quality Assessment Service (UK NEQAS) helps to ensure t
Connection: The American Society of Clinical Oncology (ASCO) and the College of American Pathologist
would recommend that standardization Ibegin with identifying a reliable and trustworthy source ...
Interview Immunohistochemistry for Oestrogen and Progesterone Receptors Dr. Andrew Lee Consultant H
Connection: What is the difference between the H score and the Allred score? Which is better? What d
Connection: Can you comment on the burden in the laboratory, if one changes from a current ER/PR ass
Opinion & Interview IHC Standardization: A Dako Perspective Dr. Ole Rasmussen R&D Director,
Connection: Dako has developed Readyto-Use Antibodies for in vitro diagnostic applications. How is t
Articles UK NEQAS-ICC & ISH: Historical perspective, current role, future directions Andy Dodso
UK NEQAS-ICC in the 1990s In his first year as Scheme Organiser, Keith oversaw the transition to sub
The application could be argued to represent a field change in terms of the rigour with which the an
Assessment teams consist of four assessors, who view slides around a multi-headed microscope and sco
The archive which UK NEQAS holds, both in terms of stained slides and methodological data, must sure
For Immunocytochemistry and FISH RESULT: RUN 80L SLIDE: NEQAS Laboratory No: XXX Mr. A. Scientist De
Figure 6. Feedback on results has always been given high priority, and for many years this has been
a b c d Figure 7. The antigen chosen by Gerry Reynolds for his very first assessment run was kap
Bibliography Selected UK NEQAS-ICC & ISH papers. Ibrahim M, Dodson AR, Barnett S, Fish D, Jasani
Articles Nordic Immunohistochemical Quality Control (NordiQC) An Organization for External Quality A
parameters (i.e. results interpretation and reporting) (4, 5). In an EQA setting, by circulating ser
CD79a (Fig. 2) Among 112 laboratories submitting stains in the latest run, most used mAb clone JCB11
References 1. Rhodes A, Jasani B, Barnes DM, Bobrow LG, Miller KD. Reliability of immuno-histochemic
Fig. 2. CD79a A. Optimal CD79a staining of the tonsil using the monoclonal antibody (mAb) clone JCB
Standardization of Ki-67 Immunohistochemical Staining for Diagnosing Grade of Gastrointestinal Strom
was conducted in 49 GIST cases. The concordance rate for the evaluation results at three laboratorie
CB pH6 a b c d TE pH9 e f g h Autoclave 121° C/10 min Water bath 95° C/40 min Microwave
50 Table 1. Correlation between NCC and STD methods R2=0.9483 Categories of proportion NCC Method 3
Opinions Importance of Standardization for Predictive Prognosis David J. Dabbs, MD Chief of Pathol
is documented and serves as a surrogate marker for the initial exposure to formalin. Since the first
Opinions The New Era for ER and PRIts time to Standardize! Dr. Ian Ellis, B.Med.Sci. BM, MS, FRCpa
et al 2001). The main reason for false-negative results is due to inefficient heat-induced epitope (
Standardization of HER2 TestingInconsistency Raises Questions Opinions Sunil S. Badve, MD, FRCPath
rence seen in these trials is in the order of 50%. This is the major reason for all the excitement a
which now requires expression of HER2 by at least 30% of tumor cells (instead of 10%). It has also r
IHC CONSENSUS MEETING, JANUARY 27 2008, SANTA BARBARA, CA, USA , IHC CONSENSUS MEETING, JANUARY 27,
Richard Cartun, PhD, Sunil Badve, MD Jon Askaa, PhD Søren Nielsen, HT, CT, Mogens Vyberg, MD Elizab
Dako Abstracts Abstracts presented at the 30th Annual San Antonio Breast Cancer Symposium December
Dako Abstracts Amplification of ESR1 may predict resistance to adjuvant tamoxifen in postmenopausal
Dako Abstracts Abstracts presented at the United States and Canadian Academy of Pathology (USCAP) A
Dako Abstracts Metastatic Pancreatic Endocrine Tumors in the Liver Express KOC Briones AJ, Bourne P
Dako Abstracts Merkel Cell Carcinomas Express K Homology Domain Containing Protein Overexpressed in
Dako Abstracts Immunohistochemical Analysis of KOC/IMP3 in Malignant Pleural Mesothelioma Xu H, Sim
Dako Abstracts KOC, TTF-1 and CDX2 Discriminate Small-Cell Carcinoma from Carcinoid and Pancreatic
Dako Publications Publications Co-authored by Dako: In Press Li L, Xu H, Spaulding B, Cheng L, Si
Dako Meetings 2007 - National Society for Histotechnology Meeting. Denver, CO NSH workshop attend
New Premium Quality Concentrated Antibodies New Products Your current and future needs drive the c
Targeting Colorectal Adenocarcinomas, Anti-TIMP-1 and Anti-Villin Monoclonal Mouse Anti-Human Tissue
Dakos FLEX Ready-to-Use Concept Enhance performance of your laboratory by running FLEX Ready-to-Use
2008 Product Catalog Available Now! The catalog features more than 170 new products, including the A